We guarantee you that a rebuild pump will have the same performance and warranty as a new pump. Rebuilding an Ion Getter Pump means that we:
– open the pump and disassemble the elements.
– clean stainless steel components and replace the titanium/ceramic parts.
– vacuum degas all metal parts.
– install the elements in the clean body.
– leak check the pump.
– bake out on vacuum manifold.
– check leakage current and pinch off seal.